1. General
The MFR, UVMFR, and SPUV all use new technology "Ion Assisted Deposition" interference filters. The slit width and effective center wavelength are exceptional and stable. However, while the throughput/transmission of IAD filters is far superiror compared with previous generations of optical filter technologies, we still recommend that yearly factory calibrations.
Yes, our model UVA-1 covers the entire UV-A and UV-B range of 280-400 nm ... but since only a tiny fraction of the integrated dose involves the UV-B range of 280-320 nm (compared with the total 280-400 nm range), what you are essentially measuring is UV-A.
Ideally you should use the tables and the solar zenith angle at the time of measurement. The origin of the multiply by 25 vs. divide by 40 is that some radiometers are calibrated in watts/m^2 while others are in calibrated in microwatts/cm^2. The difference between the two is a factor of 1000 (or 25 *40). This is a direct scale to the UV-Index, which was derived by WMO in 1994. There were UVB meetings at Diableretes, Switzerland in 1994 and 1996 to agree on this.
|